Description
Medical automatic manifold system is designed to provide a continuous and uninterrupted gas supply to the facility without any manual adjustments. This system automatically switches over when the primary cylinder bank is depleted. Even in case of a power failure, the system continues to supply gas without interruption. The system is designed in accordance with NFPA 99 and ISO standards.
LED display, full-automatic manifold;
Suitable for Oxygen, Air, Nitrogen,N₂O,CO₂;
Remote alarm system;
Electric heater function is optional;
Wall or floor mount installation available


Technical Specifications:
Installation Dimensions:
Product Dimensions:
Specifications
Input pressure: 4-200bar
Output pressure: 3-10bar(Adjustable)
Input power: AC110-240V, 50/60Hz
Working Voltage/Current: DC24V, 250mA
Maximum output flow: 100m³/h
Changeover Pressure: 6-10bar(Adjustable)
Changeover time: 3S
Alarm Signal: Sound and light simultaneously ;
Ambient temperature:-20~40℃;
Ambient humidity: ≤85%;
Pressure units: MPA, PSI, KPA, Bar